             |
The pressure
gauge has been calibrated to plus or minus ¼ psi (mid range) and
the thermometer tested to plus or minus ¼ degree prior to shipment
of the instrument. If after a period of use there is a question
of gauge and thermometer accuracy, the following checks should be
performed on the unit:
- The gauge
should be checked and calibrated using the Zahm & Nagel tester
(series 8000), a "dead weight" tester or a digital test gauge.
The most accurate range of the gauge is the midrange, and the
gauge should be calibrated to this area. The needle can be re-set
by removing the face cover and inserting a small screwdriver into
the slotted brass screw located at the 5 o'clock position of the
gauge. Turn the screwdriver clockwise or counterclockwise to match
the pressure established on the gauge tester or dead weight tester.
A "zero zone" is located on the gauge face and the needle may
occupy any place in this zone and still be in calibration-do not
attempt to calibrate the gauge back to this zone.
- The thermometer
liquid column may become separated due to rough handling and may
be rejoined as per enclosed instruction sheet-"separated columns."
If the thermometer is broken please refer to replacement instructions
sent with the instrument.
- The piston
(1005) may become stuck in the piston body (1006)-if this condition
occurs, unlock the piston locking pin (1008) open the inlet valve
(1020-B) and pull the piston out to full open position. Apply
a small amount of "o" ring lubricant to the piston and work the
piston up and down to lubricate the internal "o" ring (1009) located
inside the piston body. This should be done at frequent intervals
as part of the routine maintenance of the instrument.
Warning:
Operator must wear eye protection while operating this instrument.
Click Here for the line drawing and parts
description that show the components of the volume meter. Refer
to this page when operating the instrument or when ordering parts.
TO
OPERATE:
Note: Before
operating, cool the volume meter to the approximate temperature
of the beer being tested .
- Close the
lower inlet valve (1020-B) and lock the expansion piston (1005)
in the "in" position.
- Open the
top vent valve (1020-B) and establish counter pressure with aspirator
bulb (1042) (or other convenient pressure source) up to the approximate
counter pressure plus hydrostatic head pressure existing in the
tank to be sampled and then close the top vent valve. (1020-B)
- Attach the
volume meter to the trycock of the tank by opening the clamping
device screw (1024) and sliding the clamping device yoke (1023)
over the boss of the trycock. Turn the clamping device screw clockwise
so that the clamping device gasket (1025) makes a tight seal between
the try cock and clamping device assembly(1021).
- Open the
inlet valve (1020-B) and then fully open the trycock.
- When the
gauge pressure is steady, remove the aspirator bulb (1042) and
partially open the top vent valve (1020-B) so that the pressure
on the gauge drops about 1 psi. Allow the volume meter to fill
until beer overflows at the gooseneck(1045).
- Fully open
the vent valve (1020-B) so as to allow the beer flow to sweep
out any air bubbles
- Close the
inlet valve, vent valve and the trycock. Remove the volume meter
from the trycock on the tank.
- Partially
open the vent valve (1020-B) to reduce the false pressure reading
and allow the gauge pressure to drop to approximately the equilibrium
pressure expected in the test, and then close the vent valve (1020-B)
- Turn the
piston handle (1007) counterclockwise to the release position.
Do not pull the handle out but allow the bottle pressure to force
the piston out as the sample is shaken. Pulling the handle out
will result in damage to the gauge.
- Vigorously
shake the volume meter until equilibrium temperature and pressure
is established and record the readings. Refer to the chart accompanying
the instrument for the corresponding gas volumes. Equilibrium
is achieved when, after shaking, the temperature and pressure
remain constant. Do not allow your hands to touch the metal body
of the volume meter when shaking as this will cause a rise in
temperature of the beer. Hold the volume meter by the handles-if
the volume meter is not equipped with handles hold it by the base
shield (1003) and the neck shield (1004a).
Note: the volume
meter chart (1049) gives solubility of Co2 gas in average beer,
and will be found to give lower readings than charts based on solubility
in water.
|