|
Care of the
Instrument:
The Series 14000
computerized piercing device is a precision instrument with a microprocessor
and requires daily maintenance to assure that accuracy is maintained.
After finishing
tests, remove the computer (14003) from the piercing device, clean
the pressure port with a soft cloth and make sure there is no dirt
or restrictions in the port.
Store the computer
in a dry place.
Clean the temperature
sensor’s probe with a wet cloth to remove product buildup. Do not
drop or bend probe. Store in a safe place.
Backflush the
piercing device with warm water by attaching a hose from a water
source to the nipple on the snift valve. Allow the water to run
for several minutes to flush out the system.
Rinse all outside
surfaces of the piercing device with warm water and dry with a soft
towel. Store in a dry place.
Preparation of the
Instrument Prior to Measurement:
Connection
of the Temperature Sensor (14004):
Insert the connector
of the temperature sensor (14004) into the temperature sensor connecting
port of the 14003 computer. Align the four pins located in the end
of the connector cord with the connector port located on the top
of the 14003 computer. Insert into port and tighten metal ferrule.
To insert the probe of the temperature sensor into the piercing
device, loosen the adapter nut (6009) on the thermoadapter (6007)
and push probe into the adapter. Now tighten adapter nut enough
to allow the probe to move up and down freely.
Installation
of Batteries:
Remove the battery
holder cover (located at the rear of the 14003 computer) by turning
the retaining screw clockwise until loose. Remove the cover and
install two "AA" alkaline batteries in the battery compartment.
Be sure to check the polarity before installing the batteries.
Connection
of 14003 Computer to the Piercing Device:
Insert 1/8 NPT
male pressure port fitting of the 14003 computer into the gauge
adapter (5039) of the piercing device. Tighten the gauge adapter
nut with a wrench until the 1/8 NPT fitting seats firmly against
the gauge gasket (5083).
Power On/Off:
To turn power
ON, press the ENTRY key. "ON" appears on the display and
0.0 pressure reading is displayed.
To turn power
OFF, press ENTRY key and hold down for three seconds. The power
will also turn off automatically when no pressure change has been
detected for three minutes.
Setting the
Display Mode:
By pressing
the ENTRY key, the display mode changes from pressure, regular gas
volume, diet gas volume, temperature, and then pressure in that
order. The bar at the right side of the display will show the present
mode.
- Pressure
(No bar)

- Regular Gas
Volume (Upper right side)

- Diet Gas
Volume (Middle right side)

- Temperature
(Lower right side)
The following set
of temperature/pressure units are available:
- Pressure
= kgf/cm2, Temperature = C.
- Pressure
= psi, Temperature = F.
Setting the
Units:
When you remove
the battery cover, you will see dip switches in the lower left of
the compartment. Turn off the power and set the dip switches as
below:
- Pressure
- kgf/cm2, Temperature = C.

- Pressure
= psi, Temperature = F
Altitude
Input:
By inputting
an altitude adjustment, the measured gas volume is adjusted correctly.
The following steps will allow the operator to perform these adjustments.
- Press UP
key while simultaneously pressing CAL key. A blinking "0"
is displayed.
- Press UP
or DOWN key to display the value that indicates the altitude where
the test is being performed. If the altitude is 0 m, set the display
to read 0. If the altitude is 100 m, set the display to read 10.
If altitude is 1200 m, set display to 120. If altitude is 3000
m, set display to read 300. Set the altitude in increments of
100 m. The factory setting is 0 m.
- Press ENTRY
key to complete the setting. To cancel setting, press CAL key.
If batteries are
exhausted, no setting will be possible.
Zero Setting
and Calibration:
PRESSURE ZERO-SETTING
If no pressure is applied to the instrument and a
displayed pressure of 0.00 psi is not shown, then a zero-setting
should be performed as follows:
- Set the display
pressure mode to "PRESSURE".
- Confirm that
the instrument is not subject to an outside pressure source.
- Press CAL
key. "000" blinks twice in the display. Zero-setting
is complete.
Note: Zero-setting
is not available when pressure and temperature are displaying "error".
Calibration
of Pressure Span:
Calibration
of pressure span is performed to adjust the displayed pressure value.
Adjustment is available between 3 and 5 kgf/cm2 or 42.8 and 70 psi.
Never subject this instrument to pressures higher than 5 kgf/cm2
or 70 psi. To calibrate the pressure:
- Turn instrument
on.
- Set the display
mode to "PRESSURE".
- Perform the
zero-setting.
- Attach a
master pressure gauge in line with the computer. To do this, a
Zahm sample bottle is placed in the instrument and pierced. Establish
a hose connection between the master gauge and the Zahm sample
bottle. Introduce pressure on the inlet side of the master gauge
into the sample bottle. This will establish a known pressure to
the computer, and calibration adjustments can then be made.
- Press CAL
key for three seconds or more. A blinking "CAL" is displayed.
- Press ENTRY
key. The present pressure is displayed in the blinking mode.
- In case the
displayed pressure is not equal to the true pressure value, adjust
by pressing UP or DOWN keys.
- Press ENTRY
key. "END" is displayed and the display returns to Pressure
Mode.
To cancel setting
span in (6) and (7), press CAL key.
If computer
is connected to a dead weight tester, avoid introducing oil into
the pressure inlet of the computer.
If "ERROR"
is displayed, pressure span can not be calibrated.
Calibration
of Temperature:
Calibration
of temperature is performed to adjust the displayed temperature.
- Set the display
mode at Temperature C or F.
- Put the temperature
probe in a water bath with a constant controlled temperature
- Press CAL
key for three seconds or more. A blinking "CAL" is displayed.
- Press ENTRY
key. A blinking temperature is displayed.
- In case the
displayed temperature is not equal to the true temperature, adjust
it by pressing the UP or DOWN keys.
- Press ENTRY
key. “END” is displayed and display returns to Temperature mode.
To cancel setting
temperature in (4) and (5), press the CAL key.
To return calibration
to the manufacturer’s setting, press the UP and DOWN keys simultaneously
at step (5). After a blinking “DEF” is displayed twice, the display
returns to this setting.
If “ERROR” is
displayed, you will not be able to calibrate the temperature.
Data Memory:
Writing of Data:
A maximum of five measured values can be recorded.
- Press the
DOWN key while the first measured value is displayed. “ADD” is
displayed and the data is recorded in Memory (1). Display returns
to the indicated measured value.
- Press the
DOWN key while the second measured value is displayed. “ADD” is
displayed and the data is recorded in Memory (2). Display returns
to the indicated measured value.
- Repeat the
above process for Memories (3), (4) and (5).
- When the
sixth value is added, the first data is erased. The second data
moves to (5), the third to (4), the fourth to (3), the fifth to
(2) and then the sixth is recorded in Memory (1).
Data can not
be written if pressure or temperature is displayed as “EEE” or “–
–” or if batteries are exhausted.
Reading Data:
- Press UP
key in the measuring mode. After (1) is displayed, the value in
Memory (1) is displayed.
- Within five
seconds press UP key again. After (2) is displayed, the value
in Memory (2) is displayed.
- Repeat this
process for reading data in Memories (3), (4), and (5).
“– –” is displayed
when no data is in memory.
If UP key is
not pressed in five seconds, the display returns to the first measured
value.
Press CAL key
to return to the measured value.
If batteries
are exhausted, you will not be able to read data.
Clearing
Recorded Data:
- Press UP
key in the measured value display mode. Soon after (1) is displayed,
the value in Memory (1) is displayed.
- Press UP
and DOWN keys simultaneously for five seconds. All the data in
Memory (1) through (5) is cleared. A blinking “DEL” is displayed
and the display returns to normal.
Error Indication:
Blinking indication:
Batteries are exhausted. Replace with new ones.
EEE indication:
When pressure or temperature is out of range, EEE is displayed in
each mode. Also when regular gas volume or diet gas volume is out
of measuring range, EEE is displayed in each mode.
– – indication:
Gas volume is under 0.03 or no data was recorded.
|