|
Cleaning
of the Instrument
When the tests
have been completed, the instrument should be cleaned as follows:
- Close the
Whitey valve and open the burette cock and allow the caustic solution
to drain into the level bottle. Discard the caustic solution if
it has become weak or, if still good, store in a sealed glass
container for future use.
- Open the
Whitey valve, fill the level bottle with warm water and raise
the level bottle above the top of the burette. Fill the burette
with warm water until it overflows at the caustic trap.
- Lower the
level bottle and allow the water in the burette to drain through
the Whitey valve and also back into the level bottle. Repeat step
#2 several times or until all traces of caustic solution have
been drained from the burette and fittings.
- The pressure
gauge should be removed from the gauge adapter (5039), rinsed
in warm water and then dried.
Note:
Caustic solution will eventually discolor the Tygon Plastic
connection tubing (5052) and also cause deterioration of
the rubber coupling (5046). If this condition occurs, the
parts should be replaced.
Calibration
of Pressure Gauge:
The pressure
gauge should be tested for accuracy at the beginning of each shift.
Procedures for re-calibration are sent with the instrument and are
available upon request from Zahm & Nagel Co.
|