After the tests
have been completed the instrument should cleaned and checked as
follows:
- Backflush
the piercing assembly system with warm water. To do this, attach
a hose to a water supply and then to the Teflon nipple of the
Whitey valve. Open the Whitey valve and allow the water to flush
through the system for several minutes.
- Rinse the
entire instrument with warm water and dry with a soft towel.
- Check the
dial thermometer for accuracy (re-calibration instructions are
shipped with the instrument). Before installing the dial thermometer,
apply a small amount of "O" ring lubricant to the stem and work
the thermometer up and down to lubricate the internal "O" ring.
- Check the
pressure gauge for accuracy (re-calibration instructions are shipped
with the instrument). Be sure to install gauge gasket in the gauge
adapter before the pressure gauge in installed on the instrument.
Note: All replacement
parts for the 6000 series are the same as used on the Zahm New Style
Air Tester 5000 Series except for parts #6003 through #6018 as listed
on the parts list.
|