             |
When the tests
have been completed the instrument should be cleaned as follows:
- Close the
Whitey valve and open the burette cock, allowing the caustic solution
to drain into the level bottle. Discard the caustic solution if
it has become weak or, if still good, store in a sealed glass
container for future use.
- Open the
Whitey valve, fill the level bottle with warm water and raise
the level bottle above the top of the burette. Fill the burette
with the water until it overflows at the caustic trap.
- Lower the
level bottle and allow the burette to drain through the Whitey
valve and back into the level bottle. Repeat step #2 several times
or until all traces of caustic solution have been drained from
the burette and fittings.
- Remove the
pressure gauge from the gauge adapter (5039), rinse in warm water
and dry.
- Remove the
dial thermometer, rinse off the stem and dry. Apply a small amount
of "O" ring lubricant to the stem and install the thermometer
in the tester. Move the thermometer up and down several times
to lubricate the internal "O" ring of the tester.
Note:
Caustic solution will eventually discolor the Teflon plastic connection
tubing (5052) and also cause deterioration of the rubber coupling
(5046). If this condition occurs, the parts should be replaced.
Calibration
of the Pressure Gauge/Dial Thermometer:
The pressure
gauge and thermometer should be tested for accuracy at the beginning
of each shift. Procedures for recalibration are sent with the instrument
and are available upon request from Zahm & Nagel CO.
|